Portable energy dispersive X-ray fluorescence for metal determination of in natura water samples
Palavras-chave:
portable XRF, water, in natura, metalResumo
A portable EDXRF equipment was used to perform a qualitative and a quantitative in situ analysis of in natura water. Using this equipment 10 mL of water was directly analyzed. Detection limits, in mg L-1, for Cr, Mn, Fe, Co, Ni, Cu, Zn, Se, Hg and Pb were 12.5, 7.5, 4.9, 2.7, 2.8, 2.1, 1.6, 1.9, 3.4, 3.7, respectively, with deviation around 10%. In real samples, Fe, Zn and Pb were identifi ed. Given the obtained results, the analytical system performance appeared to be promising.
Downloads
Publicado
Edição
Seção
Licença
Autores mantém os direitos autorais e concedem à revista o direito de primeira publicação, com o trabalho simultaneamente licenciado sob a Creative Commons Attribution License que permitindo o compartilhamento do trabalho com reconhecimento da autoria do trabalho e publicação inicial nesta revista.
Esta revista proporciona acesso público a todo seu conteúdo, seguindo o princípio de que tornar gratuito o acesso a pesquisas gera um maior intercâmbio global de conhecimento. Tal acesso está associado a um crescimento da leitura e citação do trabalho de um autor. Para maiores informações sobre esta abordagem, visite Public Knowledge Project, projeto que desenvolveu este sistema para melhorar a qualidade acadêmica e pública da pesquisa, distribuindo o OJS assim como outros software de apoio ao sistema de publicação de acesso público a fontes acadêmicas. Os nomes e endereços de e-mail neste site serão usados exclusivamente para os propósitos da revista, não estando disponíveis para outros fins.